Every indicator, alarm and warning the product can show, what it means, what usually caused it, and what to do about it.

This page lists every indicator, alarm and warning the product can show, what it means, what usually caused it, and what to do about it.

Board indicators

The names below are the names in the design data. On the board, each LED is marked with its reference designator.

IndicatorWhat it meansLikely causeWhat to do
5V LED (D3)The 5 V rail is present—If it is off, the board is not powered or the mains supply has failed
5V LED (D38)The isolated 5 V rail is present—If it is off, the isolated analogue front end has no power; check the isolated supply
Alarm LED (D19)The alarm relay output is activeAt least one alarm is activeLook at the alarm list on the screen or in the web interface
Status LED (D20)The status relay output is activeThe unit is runningNo action
AUX1 LED (D21)Auxiliary output 1 is activeAn alarm in the Aux 1 mask is activeCheck the Aux 1 alarm mask
AUX2 LED (D22)Auxiliary output 2 is activeAn alarm in the Aux 2 mask is activeCheck the Aux 2 alarm mask
Klep1–Klep4, Klep5 LEDsThat output is being drivenThe unit is humidifying, flushing, or the output is stuckSee the valve section of the web interface
Motor LED (D45)The motor output is being drivenThe pump is runningSee the pump row in the web interface
Ethernet yellow LEDThe Ethernet link is up; it blinks with traffic—If it is off, the network cable or the switch port is the problem
Ethernet green LEDThe link runs at 100 Mbit/s—Off means a 10 Mbit/s link or line isolation

Screen indicators

IndicatorColourMeaning
Pumpgreen when running, grey when stoppedThe pump is running
K1–K4green when open, grey when closedThe corresponding atomisation valve is open
Alarmred when active, grey when notAt least one alarm is active
Statusgreen when active, grey when notThe unit is running
The alarms buttonred when an alarm is active, green when there is noneOpens the alarms screen
Mode: Standby / Mode: Automatic / Mode: TestThe operating mode
Flushing...A flush routine is running
Test function in progressThe unit is in test mode
Check device connectionThe screen cannot reach the controller over its RS-485 link
Failed to read alarm historyThe controller did not answer the history request
No alarm history availableThe controller has no stored alarm history

Screen banners

The banner under the mode panel explains why the unit is not humidifying.

BannerMeaningWhat to do
Output reduced due to Highlimit transmitterThe air leaving the unit is already too humid, so the output is being cut backCheck the humidity in the supply air. If it persists, check the high-limit transmitter
Waiting: AirflowNo airflow signalCheck the wiring at J14 (Airflow) and that the fan is running
Waiting: HygrostatNo hygrostat signalCheck the wiring at J16 (HL_ONOFF) and the hygrostat itself
Waiting: MasterThe master enable input is not activeCheck the wiring at J12 (MEn) and the master switch
Waiting: Alarm <name>An alarm is blocking the runAcknowledge the alarm, see the table below
Test function in progressThe unit is in test modeIt returns to Standby when the test duration is over

Web interface indicators

IndicatorMeaning
System Status: Running / StandbyThe unit is following the demand / it is not
Valve controller: running / stoppedThe demand controller is active / it is held off
Motor temperature: Overheated / NormalThe motor thermistor has tripped
Master enable: On / OffThe master enable input is active
Hygrostat: Closed / OpenThe hygrostat contact is closed / open
Airflow: Present / AbsentThe airflow input is active
Input SignalWhich demand source is selected: 0-10 V, 4-20 mA, On/Off, External demand register, Return humidity, 0-10 V or Return humidity, 4-20 mA
Active Alarms: 0No alarm is active

Web interface error message

MessageMeaningWhat to do
Invalid input ignored: <field names>One or more fields in the form could not be accepted. The other fields in the form were still savedCorrect the listed fields and press Save again

Alarms

The names below are the names the screen shows. Each one is either transient — it describes a condition that is happening right now and clears by itself — or sticky — it stays until you acknowledge it.

AlarmKindWhat it meansLikely causeWhat to do
Water Temperature HighstickyThe water temperature stayed above the threshold for long enoughHot water supply, or a hot ambient temperatureFix the water supply, then reset the alarm
Water Temperature Too HightransientThe water temperature is above the threshold right nowHot water supply, or a hot ambient temperatureCheck the water supply. The alarm clears by itself when the temperature drops
Conductivity HightransientThe conductivity is above the threshold right nowConcentrated or dirty waterCheck the water supply and the flushing interval
Conductivity After FlushstickyThe conductivity stayed above the threshold for long enough, or flapped too oftenThe water itself is too conductive; flushing cannot fix itChange or treat the water supply, then reset the alarm
Conductivity Sensor FaulttransientThe conductivity probe is not giving usable readingsThe probe is disconnected, unpowered, or the probe wiring is wrongCheck the probe at J37 and the isolated supply
Water Pressure LowstickyThe inlet pressure stayed too low for long enoughA closed valve, a blocked filter, a dry tank, or a leak in the water supplyCheck the water supply between the tank and the pump
Inlet Pressure Too LowtransientThe inlet pressure is too low right nowSame as aboveSame as above
Water Pressure HighstickyThe pressure after the pump stayed too low for long enoughA blocked filter, an air lock, or a failing pumpCheck the filter and the pump
Pump Pressure Too LowtransientThe pressure after the pump is too low right nowSame as aboveSame as above
Motor Temperature HightransientThe motor has tripped its own thermistorThe pump is blocked, the motor is overloaded, or the ambient temperature is too highCheck the pump and the water flow. The alarm clears by itself when the motor cools down
No Motor DrivestickyThe pump is commanded on but the motor is not drawing currentThe motor, its wiring, or the motor contactor is faultyCheck the motor and the wiring at J33
Board Temperature HightransientThe temperature inside the controller is at or above 65 °CPoor ventilation, or a hot environmentImprove the airflow around the controller
Water Leak DetectedstickyThe leak input has been triggeredWater is present, or the leak detector has trippedClose the water supply and find the leak. The alarm never clears by itself — reset it once the leak is fixed
Input Signal ErrorstickyThe overvoltage input has been triggeredThe demand input is carrying more than the allowed voltageCheck the demand signal source and its wiring
Maintenance RequiredstickyThe maintenance countdown has run out, or more than 2000 pump hours have passed since the last maintenanceThe unit has been running for a long time without serviceCarry out the maintenance, then acknowledge it (see below)

Which alarms clear by themselves

Clears automaticallyNeeds a manual reset
Water Temperature Too HighWater Temperature High
Conductivity HighConductivity After Flush
Conductivity Sensor FaultWater Pressure Low
Inlet Pressure Too LowWater Pressure High
Pump Pressure Too LowWater Leak Detected
Motor Temperature HighNo Motor Drive
Board Temperature HighInput Signal Error
Maintenance Required (clears when the maintenance is acknowledged)

Motor Temperature High, Board Temperature High and Conductivity Sensor Fault clear themselves once the underlying condition goes away. The screen does not offer a reset button for them, and neither does the web interface.

Which alarms offer a reset button

AlarmScreenWeb interface
Water Temperature Highyesyes
Conductivity After Flushyesyes
Water Pressure Lowyesyes
Water Pressure Highyesyes
Water Leak Detectedyesyes
No Motor Driveyesyes
Maintenance Requiredyes - Reset on that card means "maintenance performed"no

How an alarm becomes a permanent alarm

Transient alarmBecomes sticky afterAfter this many trips in one hour
Water Temperature Too High → Water Temperature High2 minutes of continuous excess3 trips
Conductivity High → Conductivity After Flush2 minutes of continuous excess3 trips
Inlet Pressure Too Low → Water Pressure Low30 minutes of continuous excess10 trips
Pump Pressure Too Low → Water Pressure High30 minutes of continuous excess10 trips

The sticky alarm stays until somebody acknowledges it.

Acknowledging an alarm

  1. On the screen, open the alarms screen and press Reset next to the alarm.
  2. In the web interface, press Reset in the Active Alarms card.
  3. Over the fieldbus, write a bit mask to the ALARM_RESET register.

Acknowledging also clears the trip history behind the alarm, so a sensor that is still out of range will raise a fresh alarm rather than immediately latching again.

Maintenance acknowledgement

The Maintenance Required alarm is cleared by acknowledging the maintenance itself, which also resets the pump-hours counter and restarts the one-year countdown.

  • On the screen, open the alarms screen and press Reset on the Maintenance Required card. That button means "maintenance performed".
  • Over the fieldbus, write 1 to the MAINTENANCE register.
  • The web interface does not have a button for this; use the screen or the fieldbus.